Air System Troubleshooting Guide
system is:
WE SUGGEST YOU CHECK THIS:
A. The repressurization valve could be malfunctioning. If the repressurization valve is not functioning properly, a transducer in each tower would identify the abnormal pressure condition and put the dryer into a switching failure mode. The repressurization valve would either need to be rebuilt and/or replaced.
B. The purge valve could be malfunctioning. If the purge valve is not functioning properly, a transducer in each tower would identify the abnormal pressure condition and put the dryer into a switching failure mode. The purge valve would either need to be rebuilt and/or replaced.
C. The dryer could have a defective transducer. This would identify a false abnormal pressure condition putting the dryer in a switching failure mode. The transducer would need to be replaced.