Air System Troubleshooting Guide

If your
system is:
Experiencing High Dew Point (as indicated by the dew point monitor) Providing Low Performance Experiencing Frequent Starts and Stops Shutting Down Due to High Temperature Experiencing High Amp Draw and Motor Trip Out Lag Alarm Going Off (multiplex systems) Desiccant Dryer Going into Switching Failure

WE SUGGEST YOU CHECK THIS:

A. The compressor may be running continuously for long periods of time. Many compressors are designed for intermittent operation and will overheat if run too long. This could be caused by a defective pressure switch, which is not allowing the compressor to shut off when adequate pressure is attained.

B. The temperature switch itself could be defective. Its job is to protect the compressor from excessive heat by shutting it down. If the switch is providing a false temperature reading, it could be shutting down the compressor prematurely. In such cases, the switch may need to be replaced.

C. The compressor may have worn rings or intake/discharge valves. As mentioned above, if these components are worn, the compressor will not create adequate pressure, resulting in long run times and excessive heat.

Questions About Our Products?
Contact Us or Call 1-866-549-6446
Products
Industrial Vacuum & Air
Medical Gas Pipeline
Medical Devices
Legacy
Services
Industrial Air System Repair & Maintenance
Industrial Vacuum System Repair & Maintenance
Medical Gas Pipeline Preventative Maintenance
Exchange for Change Program
Custom Vacuum Pumps & Services
Troubleshooting Guides
Login
About Us
Company Profile
Applications
Contact Us
Request a Quote: vacuum & air
Request a Quote: medical gas pipeline
Squire-Cogswell
1111 Lakeside Drive, Gurnee, IL 60031 USA  |  Call 1-866-549-6446
Copyright © 2021| Legal| Terms and Conditions| Connect With Us!